Reliability, quality and convenience make the RD/ Hitachi 912 the optimal clinical chemistry workhorse system for the small up to medium workload laboratory. Sample selective analyser designed for mid-size workloads. Flexible system for the performance of both ical and special clinical chemistries, homogeneous immuno- assays, and STAT analyses. Capacity of 35 tests on board (including 3 ISE); up to 360 tests per hour, up to 720 tests with optional ISE.
Assay Types Hitachi 912
• Monochromatic • Bi-chromatic • Endpoint • Kinetic • Simultaneous endpoint and kinetic • Endpoint with sample blanking • Kinetic with sample blanking • Simultaneous double endpoint and double kinetic
Sampling System Hitachi 912
Photometric 360 tests/hour Photometric & ISE 760 test s/hour Sample volume per test: 3 - 50ul Sampling rate: Once every 10 seconds for photometric chemistries, Once every 20 seconds for ISE Bar Code Reader Formats: ...Coda bar, Interleaf 2 of 5, Code 39, Code 128
ISE System Hitachi 912
Sample volume: 15 µL Photometric & ISE 760 tests/hour
Roche/Hitachi 912 System Features
Method capacity: 68 programmable chemistries, 46 parameter files, 32 methods on-board plus 3 ISEs, reagent bottles in 20 ml, 50 ml, and 100 ml sizes. Operator interaction by touch-screen, mouse or keyboard. Online, software stores 10,000 patient results and 5,000 QC tests. Intuitive software with contest-sensitive help. Processes up to 720 tests per hour from a test menu of greater than 100 assays. Consolidate multiple sample types including serum, plasma, urine and spinal fluid. Bar-coded Reagents: automatic reagent registration, automatic bottle changeover, automatic calibration tracking, random reagent placement.
Approval Worldwide Dimensions Analytical unit: 39.6" x 30" x 45.6"
Throughput 360 photometric tests per hour; up to 720 ISE tests per hour
Test Channels 68 programmable chemistries; 46 parameter files, plus 3 ISEs (Na+, K+ and Cl-), 8 calculated tests, 8 compensated tests, 3 serum indices
Product Description
Standardized technology. Reagents, calibrators and controls interchangeable with other Roche/Hitachi analyzers to minimize inventory, purchasing, and training costs. Consistent patient results throughout your health care system. Automated sample handling. 70 positions for routine, STAT and QC samples. 2-35 ml sample volumes. Intermix sample types. Intermix 5 ml, 7 ml, 10 ml primary sample tubes and sample cups. No adapters needed .Refrigerated inner ring for up to 20 standards and controls
Our lab uses this machine for the following tests>>
- CSF Protein
- Angiotensin Converting Enzyme
- Alpha 1 Antitrypsin
- Anti thrombin III
- Fructosamine
- IgA
- IgG
- Lipoprotein A
- Urine Albumin
|